2.7 Timing Generator (TiGer)

The LEMO connectors of all TDC channels, the TRG channel, and the GATE channel can be used as an AC-coupled trigger output. The TiGer functionality can be configured independently for each connector.

Each TiGer is configured using the ndigo6g12_tdc_tiger_block struct. The tiger blocks can be triggered by any combination of inputs, including the auto-trigger and the ADC channels.

Note

The TiGer configuration is similiar to the gating blocks.

The TiGer can be used in different output modes. For an overview of the different modes, see the documentation in the API section.

With restrictions, the respective LEMO connectors can be used simultaneously as a TiGer output and as an input.